calibration of device - non-VOC 35

e-finder

TST 40 (1)

1. Calibration

  • Every 24 hours, the device needs to be calibrated.
  • Or when requested by device.
  • To calibrate, select the icon.
TST 40 (3)

2. Calibration selected

  • Select the next icon (see picture) to start the calibration
TST 40 (2)

3. Calibrate black trap

  • 1) Remove the hole plug
  • 2) Snap the spectro on the black trap to start black calibration
  • 3) Make sure the black trap is correctly positioned (see picture)
  • 4) Wait until 4 measurements are performed
TST 40 (3)

4. Calibration white plaque

  • Select next on the device.
TST 40 (5)

5. Calibration white plaque

  • Important!! White plaque needs to be positioned with hiding cover away from you. (see picture)
  • Position the device on the white plaque until the 3 led lights are green and keep it stable.
  • Device is now ready to use for 24 hours.