calibration of device - non-VOC 35 e-finder 1. Calibration Every 24 hours, the device needs to be calibrated. Or when requested by device. To calibrate, select the icon. 2. Calibration selected Select the next icon (see picture) to start the calibration 3. Calibrate black trap 1) Remove the hole plug 2) Snap the spectro on the black trap to start black calibration 3) Make sure the black trap is correctly positioned (see picture) 4) Wait until 4 measurements are performed 4. Calibration white plaque Select next on the device. 5. Calibration white plaque Important!! White plaque needs to be positioned with hiding cover away from you. (see picture) Position the device on the white plaque until the 3 led lights are green and keep it stable. Device is now ready to use for 24 hours. Repair Steps Basecoat/ Topcoat Clearcoat Primer/ Filler Body Filler Sanding Filler Wet-on-Wet Filler Start working with baslac today. It just makes sense. Let’s chat Get in touch with your local baslac dealer now and start working smarter. Find a dealer near you