设备校准 - non-VOC 35系列

e-finder 电子查色工具

TST 40 (1)

1. Calibration 校准

  • Every 24 hours, the device needs to be calibrated.

    该设备每 24 小时需要校准一次。

  • Or when requested by device.

    或当设备提示需要校准时进行校准。

  • To calibrate, select the icon.

    要进行校准,请选择该图标。

TST 40 (3)

2. Calibration selected 已选择校准

  • Select the next icon (see picture) to start the calibration 选择下一个图标(见图) 以开始校准
TST 40 (2)

3. Calibrate black trap 黑色校准块

  • 1) Remove the hole plug 取下孔塞
  • 2) Snap the spectro on the black trap to start black calibration 将测色仪卡扣再黑色校准块上以开始校准
  • 3) Make sure the black trap is correctly positioned (see picture) 确保黑色校准快已正确定位(见图)
  • 4) Wait until 4 measurements are performed 等待完成四次测量
TST 40 (3)

4. Calibration white plaque 白色校准

  • Select next on the device. 在设备上选择“下一个”
TST 40 (5)

5. Calibration white plaque 白色校准

  • Important!! White plaque needs to be positioned with hiding cover away from you. (see picture) 重要!!

    白色校准板必须使遮光盖远离你(见图)。

  • Position the device on the white plaque until the 3 led lights are green and keep it stable. 将设备放置在白色校准板上,直到 3 个指示灯均亮绿,并保持稳定。
  • Device is now ready to use for 24 hours. 设备现在可正常使用24小时。